Flow Rate Accuracy: Within ±3% (after correct calibration)
Mechanical precision: Within ±2%
Bolus Rate: 50/60ml Syringe: 1,200ml/h
30 ml Syringe: 720ml/h
20 ml Syringe: 480ml/h
10 ml Syringe: 240ml/h
Purge Rate: 50/60ml Syringe: 1,500ml/h
30ml Syringe: 900ml/h
20ml Syringe: 600ml/h
10ml Syringe: 300ml/h
Volume Limit: 0.1ml-9999.9ml (0.1ml step)
Total Injection Volume: 0.1ml-9999.9ml (0.1ml step)
Occlusion of Single Channel Syringe Pump:
High: 800mmHg ±200mmHg (106.7kPa±26.7kPa)
Medium: 50/600mmHg ±100mmHg (66.7kPa±13.3kPa)
Low: 300mmHg ±100mmHg (40.7kPa± 13.3kPa )
Alarms: Injection soon finish, end of injection, occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc.
Power Source: AC 100V-240V, 50/60Hz;
Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup
KVO rate: 1ml/h
Fuse: F1AL/250/60V, 2pcs inside
Power Consumption: 30VA
IP Classification: IPX4
Equipment Classification: Class II, internal power supply, Type CF
Operating Condition:
Ambient temperature: +5℃——+40℃
Transport & Storage Condition:
Ambient temperature:-30℃——+55℃
Relative humidity: ≤95%
Dimension: 280mm (L) ×210mm (W) ×130mm(H)
Contact Information: Live Support: Chat with us online Email: service.alandental@gmail.com If you have any problem, please do not hesitate to contact us.
Shipment Note: If your package shipped by Express Delivery, please give us a valid delivery phone number, and also a full and accurate shipping address.